Error Message:
"LO1 (or LO2) not in quadrature: <list Vdc values>. Increase quadrature
specification and try again"
Description
:
Applies to additive measurements with mechanical phase shifters only. Occurs if
quadrature (90° phase shift) is not achieved within the tolerance of the quadrature specification.
Troubleshooting
:
Check that appropriate phase shifters are being used for the given carrier
frequency. Check that the Phase Shifter setting is set appropriately to either 'Mechanical'.
Adjust cable lengths so that the time delay in the DUT path and LO paths is as closely matched
as possible.
Check the Vdc reading with the DC Monitor tool and adjust the phase shift until the Vdc reading
is as close to 0Vdc as possible. Check that the Vdc reading is as close to 0Vdc as possible
before each acquisition attempt and increase the quadrature specification until a measurement
initializes.
Check quality of RF cables.
Error Message:
No zero crossing detected on LO1 (or LO2) path
Description
:
Applies to additive measurements with HX5100's only. The instrument is not able
to tune the HX5100's to a 90° phase shift relative to the DUT path.
Troubleshooting
:
Check that the appropriate HX5100's are in use for the given carrier
frequency. Check that the tune voltage is present at the tune voltage outputs for LO1/LO2 and
that this is connected to each HX5100. Inspect quality of tune voltage cables and RF cables.
Error Message: "
Quadrature not maintained on LO1 (or LO2) path: <list vdc values. Open up
quadrature specification and try again"
Description
:
Applies to additive measurements with HX5100's only. This means that there is an
instability in the system causing the point of quadrature to move.
Troubleshooting
:
This may be caused by the DUT if this occurs with the DUT in the system,
but not during the reference (noise floor) measurement where the measurement is taken with
the DUT bypassed/removed from the system. The quadrature specification should be increased
until a measurement can be taken.