3.3 Measurement
3 - 10
Fig. 3-13 Autozero Execution
(3) Sample
measurement
Place the sipping nozzle in the sample, push the sipping lever to
sip the sample, and execute the sample measurement.
NOTE:
Pressing the [START] button on the measurement screen
without pushing the sipping lever permits measuring the
sample without sipping.
(4) After the completion of the sample measurement, store or print
the measurement data.
3.3.5 Monitor Measurement
Given below is the general measurement procedure for monitor
measurement.
(1) Switching to measurement screen
After setting the conditions, press Measurement button on the
screen to move to the measurement screen.
(2) Sipping of autozero sample
Press
[autozero] icon on the screen. Place the sipping
nozzle in the sample solution for autozero, push the sipping
lever to sip the solution, and measure the autozero. Pushing the
sipping lever carries out the process automatically until the
execution of the autozero.
NOTE:
Pressing the [OK] button without pushing the sipping lever
permits executing the autozero processing without sipping.