3.3.4
3 - 9
(2) Sample
measurement
After the completion of the baseline correction, place the sipping
nozzle in the sample, push the sipping lever to sip the sample,
and execute the sample measurement.
NOTE:
Pressing the [START] button on the measurement screen
without pushing the sipping lever permits measuring the
sample without sipping.
(3) After the completion of the sample measurement, store or print
the measurement data.
3.3.4 Time Scan
Given below is the general measurement procedure for time scan.
(1) Switching to measurement screen
After setting the conditions, press the [Measurement] button on
the screen to move to the measurement screen.
Fig. 3-12 Screen to Proceed to Measurement Screen
(2) Sipping of autozero sample
After moving to the measurement screen, the autozero
execution window is displayed. Place the sipping nozzle in the
sample solution for autozero, and push the sipping lever to sip
the solution, to measure the autozero automatically.
To execute an autozero between an interval of measurements of
sample solution, press
[autozero] icon and follow the
guidance to execute the autozero.
NOTE:
Pressing the [OK] button on the screen without pushing the
sipping lever permits executing the autozero processing
without sipping.