49
Checking Measurement Results
Pin contact
Result
Indicates FAIL if there was a contact error or PASS if there was no
contact error.
Error count
Indicates the number of probes exhibiting a contact error.
Potential meas. reliability
Consistency
Indicates PASS if the relative magnitudes of the potential
measurement results suggest the proper potential gradient near the
center. Indicates FAIL if the proper relationship is not observed, in
which case operation will not proceed to analysis.
Error rate [%]
Indicates how many potential measurement results generated
measurement errors as a percentage of all potential measurement
results. (This value is not the same as the number of probes
generating contact errors.)
Variation [%]
If multiple potential measurements were performed, indicates the
amount of variation among potential results (i.e., the coefficient of
variability) as a percentage.
Analysis results
Iteration number
Indicates the number of iterative analyses performed, calculated as
the number of iterations divided by the limit iteration number.
Coincidence
Indicates the extent to which the measured potential distribution
matches the calculated potential distribution. A value of 1.000
indicates a perfect match, while smaller values indicate a less perfect
match. (Negative values are possible.)
History
Displays the measurement history. New results are appended to
the history as they are obtained. When analysis is performed, the
history displays the composite layer resistivity, interface resistance,
and temperature. If the operating mode is set to [Meas.], only the
temperature [°C] will be shown.
3
Basic Measurement
Содержание RM2610
Страница 2: ......
Страница 12: ...8 Operating Precautions...
Страница 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Страница 54: ...50 Checking Measurement Results...
Страница 72: ...68 Setting Administrator Mode...
Страница 110: ...106 Under what circumstances do contact errors occur...
Страница 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Страница 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Страница 134: ...130 Index...
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......