80
RM2612 Resistance Calculation Software
Entry of analysis conditions
•
Upper limit on iterative calculations: 1 to 999 calculations
Initial value: 30 [calculations]
Input format: 3-digit integer “000”
•
Reduction factor: 1.0E-4 to 1
Initial value: 1.0
Input format: Decimal, up to 4 decimal places, “0.0000”
•
Resistance value initial value: Automatic calculation / user-entered value / reference
to analysis results
Composite layer volume resistivity and interface resistance input range: 1.0E-10 to
1.0E+5
Input format: Decimal, exponential
•
Finite volume model area: NORMAL, MEDIUM, WIDE
Initial value: NORMAL
•
Element size: NORMAL, FINE, SUPER FINE
Initial value: NORMAL
User-entered information
about measurement
targets
Information necessary for analysis
•
Composite layer thickness [μm]
Valid setting range: 1.0E-4 μm to 1000 μm
Input format: Decimal or exponential, 5 significant digits
•
Collector thickness [μm]
Valid setting range: 1.0E-4 μm to 1000 μm
Input format: Decimal or exponential, 5 significant digits
•
Collector volume resistivity [
Ω
cm]
Valid setting range: 1.0E-10 to 1.0E+5
Input format: Decimal or exponential, 5 significant digits
Initial value: Volume resistivity of aluminum
•
Display of candidate values for collector volume resistivity
(aluminum [2.7000E-6
Ω
cm] and copper [1.7000E-6
Ω
cm])
Analysis reliability display
function
Display of reliability of analysis results
•
Iteration number [iterations]
•
Agreement between measured potential and calculated potential (coefficient of
determination)
Temperature measurement
function
Measurement of the ambient temperature near the test fixture (using the Z2001
Temperature Sensor)
Results can be displayed along with measured values on the screen and output to a
file.
Progress display
Display of progress in the form of a progress bar
Separate progress bars are shown for potential measurement and analysis.
Condition save function
•
The application launches with the same conditions that were in effect when it was
last exited.
•
Measurement and analysis conditions can be saved to a file and loaded as desired.
File save function
•
Analysis results file
•
Potential file for analysis (can be used in repeated analysis)
•
2D potential distribution file
The user can specify the folder in which files are saved.
Test fixture continuity test
function
Test of probes using the probe check board (RM9004 Test Fixture inspection)
•
OPEN Inspection
•
SHORT Inspection
Identification of probes with continuity defects
Содержание RM2610
Страница 2: ......
Страница 12: ...8 Operating Precautions...
Страница 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Страница 54: ...50 Checking Measurement Results...
Страница 72: ...68 Setting Administrator Mode...
Страница 110: ...106 Under what circumstances do contact errors occur...
Страница 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Страница 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Страница 134: ...130 Index...
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......