67
Setting Administrator Mode
5
Configure the following settings:
Debug file
Enabling this setting causes intermediate reverse-analysis calculation
results to be output to a file (for debugging use) (initial value: OFF).
If you are unable to obtain satisfactory analysis results even after
adjusting the reduction factor and other initial values, Hioki may ask
you to send this file for examination (the file format is not publicly
available).
Enable RM2611 check
Enabling this setting performs an inspection of the RM2611 Electrode
Resistance Meter (initial value: OFF).
Use this setting if you encounter issues as a way to discover
potential causes. (Ordinarily, there is no need to inspect the RM2611
Electrode Resistance Meter.)
Selecting the
[Enable RM2611 check]
checkbox in administrator
mode will add
[RM2611]
to the
[Check]
on the main screen, allowing
issues with the instrument to be diagnosed. Inspection results
are output to a file, and you may be asked to send it to Hioki for
examination. (A special jig is needed in order to inspect the RM2611.)
Password
Set the password for accessing administrator mode. This setting may
be changed as necessary.
Initial value: rm2612
Measurement counter
Click
[Reset]
to reset the measurement counter to 0. The counter is
incremented after each potential measurement to provide a rough
guide as to the number of measurements that have been made with
the RM9004 Test Fixture. Its value is output to the analysis results
output file. Please reset the counter after replacing the RM9004 Test
Fixture’s probes.
4
Customizing Measurement Conditions
Содержание RM2610
Страница 2: ......
Страница 12: ...8 Operating Precautions...
Страница 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Страница 54: ...50 Checking Measurement Results...
Страница 72: ...68 Setting Administrator Mode...
Страница 110: ...106 Under what circumstances do contact errors occur...
Страница 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Страница 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Страница 134: ...130 Index...
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......