183
Specifications of Input, Output, and Measurement
Synchronization
source
U1 to U8, I1 to I8, DC (fixed at the data refresh interval for DC only)
For the PW8001-1x, the model equipped with the motor analysis option
Ext1 to Ext4:
When the input settings of the following channels are set to Speed
(pulse input) and the remainder left over when the pulse count is
divided by half the number of poles is zero.
Ext1: Ch. B, Ext2: Ch. D, Ext3: Ch. F, Ext4: Ch. H
Zph1:
When the Ch. D input setting is set to Origin (pulse input)
Zph3:
When the Ch. H input setting is set to Origin (pulse input)
Ch. B, Ch. D, Ch. F, Ch. H:
When the corresponding channel enters
[Individual input]
operating
mode
•
Can be selected for each wiring configuration. (U and I of the same channel are
measured in sync with the same synchronization source.)
•
The zero-crossing point of the waveform after passing through the zero-cross filter is
used as the reference when U or I is selected.
Effective
frequency range
of synchronization
source
DC, 0.1 Hz to 2 MHz (up to 1 MHz for the U7001)
Effective input range
of synchronization
source
1% of range to 110% of range
Zero-cross filter
Used in zero-crossing detection for voltage and current waveforms. Does not affect
measured waveforms.
Consists of a digital LPF and HPF filters. Cutoff frequencies are automatically determined
based on the settings of the upper and lower frequency limits as well as measurement
frequencies.
Measurement lower
frequency limit
Choose from among the following frequency values for each wiring configuration:
0.1 Hz, 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
Measurement upper
frequency limit
Choose from the following frequency values for each wiring configuration:
100 Hz, 500 Hz, 1 kHz, 5 kHz, 10 kHz, 50 kHz, 100 kHz, 500 kHz, 1 MHz, 2 MHz
Polarity detection
Voltage/current zero-crossing timing comparison method
Measurement item
Voltage (U), current (I), active power (P), apparent power (S), reactive power (Q),
power factor (
λ
), phase angle (
φ
), voltage frequency (fU), current frequency (fI),
efficiency (η), loss (Loss), voltage ripple factor (Urf), current ripple factor (Irf),
current integration (Ih), power integration (WP), voltage peak (Upk), current peak (Ipk)
See “10.4 Detailed Specifications of Measurement Parameters” (p. 203).
(2)
Specifications common to voltage measurement
See “10.6 U7001 2.5MS/s Input Unit” (p. 220) and “10.7 U7005 15MS/s Input Unit” (p. 225).
(3)
Specifications common to current measurement
See “10.6 U7001 2.5MS/s Input Unit” (p. 220) and “10.7 U7005 15MS/s Input Unit” (p. 225).
Specifications
HIOKI PW8001A961-00
Содержание PW8001-01
Страница 2: ...HIOKI PW8001A961 00...
Страница 6: ...Contents iv HIOKI PW8001A961 00...
Страница 56: ...50 Checking Connections HIOKI PW8001A961 00...
Страница 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Страница 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Страница 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Страница 258: ...252 Index HIOKI PW8001A961 00...
Страница 259: ...HIOKI PW8001A961 00...
Страница 260: ...HIOKI PW8001A961 00...
Страница 261: ...HIOKI PW8001A961 00...
Страница 262: ...www hioki com HIOKI PW8001A961 00...