76
Measuring Efficiency and Loss
3.5
Measuring Efficiency and Loss
The instrument can calculate the efficiency η (%) and loss (W) using the active power values and
motor power values, and display them. For example, the instrument can calculate efficiency and
loss between input and output of various power converters (e.g., inverters, power conditioners) as
well as motors, and simultaneously calculate the overall efficiency.
Efficiency and loss equations
You can set four or fewer equations (
η
1
to
η
4
and
Loss1
to
Loss4
) for each of efficiency (η) and
loss (Loss).
Display screen
[INPUT]
>
[EFFICIENCY]
1
Select the parameters of the input side
of the equation.
2
Select the parameters of the output
side of the equation.
Select the input-side measured power value
on the left and the output-side measured
power value on the right for each figure on
the screen. Up to six inputs and outputs can
be selected for each efficiency calculation
equation. Efficiency is calculated using the
sum of the six.
Pout1
Pout2
Pout3
Pout4
Most recently calculated
efficiency value
Most recently calculated
efficiency value
Output parameters
Input parameters
Pout5
Pout6
Pin1
Pin2
Pin3
Pin4
Pin5
Pin6
Input side
Pin = Pin1 + Pin2 + Pin3 + Pin4
+ Pin5 + Pin6
Output side
Pout = Pout1 + Pout2 + Pout3
+ Pout 4 + Pout5 + Pout6
η
100
×
|Pout| / |Pin|
Loss
|Pin| − |Pout|
•
Motor power (Pm) measurement is selectable only for the motor analysis-equipped model. Configure
settings using the motor input settings screen to measure the motor power (Pm).
•
See “Configuring the motor input settings” (p. 84).
•
Calculations across wiring configurations with different power ranges are performed using data for the larger
of the two power ranges.
•
Calculations across wiring configurations with different synchronization sources are performed using the
most recent data at the time of calculation.
To moderate measured value fluctuation
•
Measured values may exhibit variations when loads fluctuates severely or transiently. In this
case, reduce the data refresh interval (to 200 ms) and also use the average function’s moving
average mode.
•
When either the input or output is DC, variation in measured efficiency values can be limited by
using the same synchronization source setting for the channel used for DC measurement as for
the AC side.
HIOKI PW8001A961-00
Содержание PW8001-01
Страница 2: ...HIOKI PW8001A961 00...
Страница 6: ...Contents iv HIOKI PW8001A961 00...
Страница 56: ...50 Checking Connections HIOKI PW8001A961 00...
Страница 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Страница 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Страница 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Страница 258: ...252 Index HIOKI PW8001A961 00...
Страница 259: ...HIOKI PW8001A961 00...
Страница 260: ...HIOKI PW8001A961 00...
Страница 261: ...HIOKI PW8001A961 00...
Страница 262: ...www hioki com HIOKI PW8001A961 00...