106
Average Function
Average operation
• Average functions for all measured values except the peak value, integrated values, and
harmonic data when using a data refresh interval of a 10 ms.
The instrument displays the following values as the peak voltage and peak current values:
In exponential averaging mode, the peak values of all data that consists of the most recent
points
In moving averaging mode, the peak values of a subset that forms from the user-specified
number of the most recent values
•
It applies not only to display values but also to measured values saved on a USB flash drive,
measured values acquired using communications, and measured values output as an analog
signal.
• When a measured-value-related setting, such as the wiring mode or range, changes, average
calculation is restarted. In this case, since no average value exists until the next data refreshing,
measured values during this period will be invalid.
• When averaging and auto-ranging are used together, it may take longer than usual to stabilize
measured values on the correct value.
• Integrated measured values during average operation are calculated from measured values prior
to average operation.
• Internal average calculations continue even when the hold function is freezing on-screen
measured values.
• The peak hold function applies to measured values after average operation.
Behavior when an overload condition occurs
If an overload condition occurs in moving averaging mode, an over-range average value will be
calculated. If an overload condition occurs in exponential averaging mode, average calculations will
continue using internal calculation values.
IMPORTANT
•
Settings cannot be switched on a wiring configuration or channel basis.
•
Waveforms shown on the screen and D/A output waveforms are not affected by overload.
•
For more information about average calculation methods for different types of measured values,
see the average section in “10.5 Specifications of Equations” (p. 212).
HIOKI PW8001A961-00
Содержание PW8001-01
Страница 2: ...HIOKI PW8001A961 00...
Страница 6: ...Contents iv HIOKI PW8001A961 00...
Страница 56: ...50 Checking Connections HIOKI PW8001A961 00...
Страница 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Страница 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Страница 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Страница 258: ...252 Index HIOKI PW8001A961 00...
Страница 259: ...HIOKI PW8001A961 00...
Страница 260: ...HIOKI PW8001A961 00...
Страница 261: ...HIOKI PW8001A961 00...
Страница 262: ...www hioki com HIOKI PW8001A961 00...