Troubleshooting
Symptom
Check Items
Remedy and Reference
Measured values
vary and fail to
stabilize.
Is a charging circuit located
near the measurement
object?
Disconnect the circuit
breaker for any nearby
charging circuits. If this is
not possible, use the lowest
measured value as the
measurement result.
Measured
values acquired
through insulation
resistance
measurement or
PV
Ω
measurement
fluctuate over time.
(If using the
IR4057-50, use
the 1-minute value
display function.)
When the measuring object
connected is a capacitor:
Does the measuring object
have a large capacitance
component (capacitor)?
If it is possible to remove
the capacitor, do so.
If not, use the lowest
measured value as the
measurement result.
When the measuring
object connected is not a
capacitor:
This phenomenon
reflects the influence of
the measuring object
capacitance component.
This is not a malfunction or
error.
Select measured values
acquired one minute
after the start of the
measurement.
If a large capacitance
component included in the
measuring object causes
the fluctuation of measured
values, select measured
values after they have
stabilized.
HIOKI IR4057B961-00
Содержание IR4053
Страница 2: ...HIOKI IR4057B961 00...
Страница 4: ...HIOKI IR4057B961 00...
Страница 18: ...12 Operating Precautions HIOKI IR4057B961 00...
Страница 72: ...66 Functionality specifications HIOKI IR4057B961 00...
Страница 84: ...Appx 4 Insulation Resistance Measurements for Solar Cell Array HIOKI IR4057B961 00...
Страница 85: ...HIOKI IR4057B961 00...
Страница 86: ...HIOKI IR4057B961 00...
Страница 87: ...HIOKI IR4057B961 00...