63
Basic and Accuracy Specifications
Voltage Measurement
AC/DC automatic
detection range
AC detected at 30 V or greater (50 Hz/60 Hz).
(pulsating currents with an overlapping AC component of
30 V or more are detected as AC)
Effect of temperature
Measurement accuracy per 1°C × 0.1
(applied in the operating temperature range excluding
18°C to 28°C)
Overload protection
750 V AC (10 s), 750 V DC (10 s)
1200 V DC (10 s) (IR4053 and IR4055 only)
Display update interval
Within 1 s
Response time
Within 1.2 s
(when input voltage is changed from 0 V to 600 V)
AC V
oltage Measurement
Input resistance
100 k
Ω
or more (50 Hz/60 Hz)
Frequency range
50 Hz/60 Hz
Range configuration
Display range
(Auto range)
Maximum
displayed value
Resolution
Accuracy
420 V
(minimum displayed
value: 30.0 V)
420.0 V
0.1 V
±2.3% rdg ±8 dgt
(ranges in excess of
600 V are outside the
accuracy guarantee)
600 V
750 V
1 V
DC V
oltage Measurement
Input resistance
100 k
Ω
or more
Range configuration
Display range
(Auto range)
Maximum
displayed value
Resolution
Accuracy
4.2 V
4.200 V
0.001 V
±1.3% rdg ±4 dgt
(ranges in excess of
600 V
*
are outside the
accuracy guarantee)
42 V
42.00 V
0.01 V
420 V
420.0 V
0.1 V
IR4053, IR4055:
1000 V
IR4056, IR4057-50:
600 V
IR4053, IR4055:
1100 V
IR4056, IR4057-50:
750 V
1 V
* Over 1000 V for IR4053 and IR4055 only
4
1
2
3
5
6
7
付録
索引
HIOKI IR4057B961-00
Содержание IR4053
Страница 2: ...HIOKI IR4057B961 00...
Страница 4: ...HIOKI IR4057B961 00...
Страница 18: ...12 Operating Precautions HIOKI IR4057B961 00...
Страница 72: ...66 Functionality specifications HIOKI IR4057B961 00...
Страница 84: ...Appx 4 Insulation Resistance Measurements for Solar Cell Array HIOKI IR4057B961 00...
Страница 85: ...HIOKI IR4057B961 00...
Страница 86: ...HIOKI IR4057B961 00...
Страница 87: ...HIOKI IR4057B961 00...