34
Insulation Resistance Measurement
3.5 Insulation Resistance Measurement
The instrument is used to measure the insulation resistance to
determine the insulation performance of circuits and equipment.
Before starting a measurement, the voltage to be applied to the
measuring object needs to be selected.
WARNING
Observe the following to avoid electric shock, short
circuit, or damage to the instrument.
•
Do not attempt to measure insulation resistance
on a live conductor. Doing so could damage the
instrument or cause an accident that might result
in injury or death. Always turn off power to the
measuring object before starting.
•
When measuring insulation resistance, dangerous
voltage is applied to the measurement terminals. To
avoid electric shock, do not touch the metal part of
the test leads.
•
Do not touch the measuring object immediately after
measurement. Doing so may cause electric shock
due to a highly charged voltage.
•
Discharge the measuring object with the discharge
function of the instrument after a measurement. (p. 39)
• Insulation resistance is the ratio of applied voltage to leakage current.
Displayed value may not stabilize depending on the measuring object,
but it is not a failure of the instrument.
• Press the
MEASURE
key fully down until a live circuit indicator lights
up. If the button is not pressed down fully, a proper measurement cannot
be made.
• Turn off the rotary switch after use.
• Disconnection when measuring is recommended of any equipment
having a lower withstanding voltage than the test voltage, or equipment
or parts having an unknown withstanding voltage connected to the circuit
to be measured.
HIOKI IR4057B961-00
Содержание IR4053
Страница 2: ...HIOKI IR4057B961 00...
Страница 4: ...HIOKI IR4057B961 00...
Страница 18: ...12 Operating Precautions HIOKI IR4057B961 00...
Страница 72: ...66 Functionality specifications HIOKI IR4057B961 00...
Страница 84: ...Appx 4 Insulation Resistance Measurements for Solar Cell Array HIOKI IR4057B961 00...
Страница 85: ...HIOKI IR4057B961 00...
Страница 86: ...HIOKI IR4057B961 00...
Страница 87: ...HIOKI IR4057B961 00...