53
Wireless Communications Function (Z3210 required)
Event recording function
The event recording function logs the data when a measured value exceeds a desired threshold value, which
can be set with GENNECT Cross. For details, see the operation guide for the GENNECT Cross app (free of
charge). The number of recorded events can be checked using the instrument.
Hold down the two keys
for 1 s as described
above.
Event count display
• Up to 99 events can be recorded. If events
has reached 99, the event recording will
stops. When another event recording starts,
previously recorded data will be deleted.
• Some events with a duration time of less than
400 ms may not be accurately measured,
failing to detect them.*
*: Current frequency, voltage frequency,
capacitance (400 ms to 4000 ms,
depending on measured value),
temperature (type K thermocouple)
2000 ms
+