92
Error and Operation Display
4.2 Error and Operation Display
Error display
Description
Remedy
v.up
The instrument’s firmware is
being upgraded.
Do not remove the batteries until the upgrade
completes.
Err 001
ROM error
Program
When the error appears on the display, it is
necessary to repair the instrument. Please
contact your authorized Hioki distributor or
reseller.
Err 002
ROM error
Adjustment data
Err 004
Memory error
Err 005
ADC error
Hardware malfunction
Err 008
Z3210 communication error
The Z3210 is malfunctioning
or is not properly connected
Perform the following steps. (p. 25)
• Disconnect and then reconnect the Z3210.
• If you have another Z3210, replace the unit
in question with that one.
If the error continues to be displayed, the
instrument needs to be repaired.
Please contact your authorized Hioki
distributor or reseller.
HIOKI CM4141C961-00