DC High V Probe Mode
2.7
DC High V Probe Mode
Use of the P2000 DC High Voltage Probe (optional) allows you to measure DC voltage of up to 2000 V (CATIII
2000 V, CAT IV 1000 V), such as open voltage of solar panels.
WARNING
Do not use the P2000 to measure AC voltage.
The probe cannot accurately measure AC voltage. Improper measurement could
lead to electric shock. You can use the P2000 for DC voltage measurement only.
Do not measure voltage that exceeds 2000 V DC.
Doing so could damage the instrument and the P2000, causing bodily injury.
Use the P2000 to measure voltage that exceeds 1000 V.
Use of other probes could cause the operator to experience an electric shock.
NOTICE
Connect the instrument and the P2000 together with the strap when
using the L4943.
The cables and plugs will be subjected to stress, damaging them.