8.6 Message Reference
133
8
C
ha
pte
r 8
R
S-2
32
C
/G
P-IB
/L
AN
In
te
rfa
ce
s
Set and Query the Auto-Ranging Setting
Syntax
Command
:AUTorange
<1, 0, ON or OFF>
Query
:AUTorange?
Response
<ON or OFF>
Example
Command
:AUT ON
Note
• Attempting to enable auto-ranging when the Comparator or Memory function is
enabled results in a execution error.
• The auto-ranging setting applies to both resistance measurement and voltage
measurement.
Cancel Zero-Adjustment
Syntax
Command
:ADJust:CLEAr
Description
Command Clears zero adjustment.
Execute Zero Adjustment and Query the Result
Syntax
Query
:ADJust?
Response
<0/ 1 (NR1)>
0....... Zero adjustment succeeded
1....... Zero adjustment failed
The acceptable range of zero adjustment for both resistance
and voltage is
±
1000 dgt.
Description
Query
Queries whether zero adjustment has succeeded or failed.
Example
Query
Response
:ADJ?
0
Zero adjustment executed successfully.
Select and Query the Sampling Rate setting
Syntax
Command
:SAMPle:RATE
<EXFast/ FAST/ MEDium/ SLOW>
Query
:SAMPle:RATE?
Response
<EXFAST/ FAST/ MEDIUM/ SLOW>
Example
Command
:SAMP:RATE MED
Query
Response
:SAMP:RATE?
MEDIUM
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...