![Hioki BT3562 Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/hioki/bt3562/bt3562_instruction-manual_2137240082.webp)
4.13 Reset Function
76
The reset function can be used to re-initialize current measurement settings
(excluding saved panel data) to their factory defaults, or to re-initialize all mea
-
surement settings including saved panel data to factory defaults.
4.13 Reset Function
1
(The SHIFT indicator lights up.)
The Menu display appears.
2
The Reset display appears.
See
"1.4 Menu Display Sequence (SHIFT
3
Select the Reset method on the Sub Display.
SEt
........ Reset (initializes measurement settings other than those stored
with Panel Save)
SYS
....... System Reset (initialize all measurement settings)
4
ENTER
blinks.
5
Executes the Reset.
To cancel the settings:
SHIFT
key
(Main Display)
(Sub Display)
The current setting blinks.
(Sub Display)
When SYS (system reset) is selected
System Reset also initializes Panel Save data.
Executing either of them cannot initialize the LAN interface settings. To initial
-
ize the LAN interface settings, execute the reset from the
SEt
item of the LAN
settings. (p.105)
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...