Appendix 5 Configuration and Extension of the Test Leads
A
7
App
end
ix
The test lead extension is normally performed by Hioki. If you want extension
performed, contact your dealer or Hioki representative.
Observe the following points when extending test leads:
• Use the thickest lead available. Extend the lead only by the necessary
amount.
• Maintain the AC four-terminal configuration while extending the lead. Chang
-
ing the four-terminal configuration to a two-terminal configuration can result in
measurement data being affected by lead resistance and/or contact resis
-
tance, resulting in inaccurate measurement.
• Make the branch section as short as possible. Try to extend the thick lead
instead.
• Make sure the lead is insulated.
• While measuring, avoid as much as possible pulling or repositioning the test
leads after executing zero adjustment.
• Extending test leads may result in excessive voltage drop. The total resis
-
tance of the test leads and contacts must remain below 20
.
• To prevent eddy currents from affecting measurement, keep test leads away
from metallic parts.
• After extending the test leads, confirm proper measurement operation and
accuracy.
Reducing Induced Voltage
Since the instrument measures a minute resistance with AC power, it is affected
by induced voltage. Induced voltage refers to voltage that allows the current gen
-
erated in the instrument to build an inductive coupling in a lead and affect signal
lines. Since the phase of the induced voltage is shifted from that of the AC cur
-
rent (reference signal) by 90 degrees, it can be eliminated with the synchronous
detection circuit if the voltage is low. But for high levels, the induced voltage dis
-
torts the signals, causing incorrect synchronous detection. The instrument moni
-
tors induced voltage internally and generates an abnormal measurement signal
if the level rises above a certain level. Reducing the length of the lead will lower
induced voltage. Reducing the length of the branched section is particularly
effective.
Appendix 5 Configuration and Extension of the
Test Leads
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...