![Hioki BT3562 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/hioki/bt3562/bt3562_instruction-manual_2137240035.webp)
3.2 Basic Measurement Example
29
3
C
ha
pte
r 3
M
ea
su
re
m
en
t
Measurement
12
Connect the test leads to a battery.
Open-terminal voltages for the instrument are as follows:
3 m
and 30 m
ranges: 25 V peak
300 m
range: 7 V peak
3
to 3000
range: 4 V peak
These voltages derive from the load associated with charging the 1.2
F
capacitor inside the instrument.
• The open-terminal voltage for the 3 m
, 30 m
, and 300 m
ranges peaks at
4 V approximately 500 ms after the terminal is placed in the open state.
• When building a measurement line using scanners, use a relay with a dielec
-
tric strength that is greater than or equal to the open-terminal voltage for the
range being used.
13
Read the measured resistance and voltage.
See
"3.7 Displaying Measurement Results" (p.38).
See
Measured Resistance
Measured Voltage
Please refer to "Before returning for repair." of "Measured value is unstable."
(p.190) as a measurement and attention.
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...