4.2 Trigger Function
60
Specify the delay from the moment a trigger is applied to the start of measurement. By using this function,
even when a trigger is applied immediately after connecting a test object, the start of measurement can be
delayed to allow sufficient time for the measurement value to stabilize. Trigger delay can be set with 1 ms res
-
olution from 0.000 to 9.999 seconds.
Trigger Delay Settings
1
(The SHIFT indicator lights up.)
The Trigger Delay setting display appears.
2
Select
ON
.
3
The numerals indicating the trigger delay blink.
4
Set the trigger delay.
5
Applies setting and returns to the Measurement display.
To cancel the settings:
SHIFT
key
(Main Display)
(Sub Display)
The current setting blinks.
(Sub Display)
(Sub Display)
Or numeric keypads
Disabling the Trigger Delay Function
1
(The SHIFT indicator lights up.)
The Trigger Delay setting display appears.
2
Select
OFF
.
3
The Trigger Delay is disabled.
(Sub Display)
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...