8.6 Message Reference
146
Query the Memory Data Count
Syntax
Query
:MEMory:COUNt?
Response
<Memory data count>
<Memory data count>
= 0 to 400
(NR1)
Example
Query
:MEM:COUN?
Response
5
Query (Download) Memory Data
Syntax
Query
:MEMory:DATA? [STEP]
Response
<Memory data No. (NR1)>,<Measured resistance (NR3)>,<Measured
voltage (NR3)>
Memory data values are returned as data objects.
If [STEP] is omitted, all memory data objects are returned continu
-
ously.
Example
Query
:MEM:DATA?
Example
Response
1, 290.60E-3, 1.3924E+0
2, 290.54E-3, 1.3924E+0
3, 290.50E-3, 1.3923E+0
4, 290.43E-3, 1.3923E+0
5, 290.34E-3, 1.3924E+0
END
Query
:MEM:DATA? STEP
Response
1, 290.60E-3, 1.3924E+0
N
(Sent from PC)
2, 290.54E-3, 1.3924E+0
N
(Sent from PC)
3, 290.50E-3, 1.3923E+0
N
(Sent from PC)
4, 290.43E-3, 1.3923E+0
N
(Sent from PC)
5, 290.34E-3, 1.3924E+0
N
(Sent from PC)
END
Note
• Stored memory data objects are returned continuously, or one data object at a
time. The “END” character is returned as the last data object. When the “STEP”
parameter is specified, one data object is returned at a time. Sending “N” to the
instrument after receiving the data causes the next data object to be returned.
The memory index is an unsigned three-digit integer. Refer to “Measurement
Value Formats” for format details of returned measurement values.
• A terminator is appended to the end of each returned memory data object.
When sending “N” from the PC or other device, a terminator is required.
See
" Message Terminators" (p.110).
• Measured values are stored in memory when pressing the
TRIG
key, applying a
signal to the TRIG EXT I/O connector or sending the
TRG
command (while
the Memory function is enabled). Up to 400 data objects can be stored. When
the memory is full, additional measurement data is not stored.
• When the Memory function is enabled, auto-ranging is disabled.
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...