86
Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Troubleshooting
022
1.
Make sure the correct patient circuit is set
up for the required therapy. See page 16.
2.
Make sure a tight seal is maintained on the
face mask or mouthpiece.
3.
Examine the connections between the
patient circuit components, tubings, and
the control unit for any source of leakage. If
necessary, disassemble the patient circuit
and assemble it again. If you see cracks on
the patient circuit components, replace the
parts. See “Replacement Parts/Kits” on page
72.
4.
Press
return
after you read the on-screen
message.
023
1.
Make sure only the Hill-Rom approved bio-
filter (206680) is used. For instructions on
how to connect the filter correctly, see page
16.
2.
Press
return
after you read the on-screen
message.
024
1.
Make sure a tight seal is maintained on the
face mask or mouthpiece.
2.
Examine the connections between the
patient circuit components, tubings, and
the control unit for any source of leakage. If
necessary, disassemble the patient circuit
and assemble it again. If you see cracks on
the patient circuit components, replace the
parts. See “Replacement Parts/Kits” on page
72.
3.
Press
return
after you read the on-screen
message.
025
1.
Make sure there are no other bio-filters near
the control unit, other than the one that is
connected to the control unit.
2.
Press
return
after you read the on-screen
message.
Error Code
Do this:
Draft_E_2019-Apr-29_Cleaned