
Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
83
Troubleshooting
Warning Messages
Warning messages show on-screen
to let you know that a critical issue
has occurred. Follow the on-screen
instructions to troubleshoot the
issue. If an error code shows, see
“Solutions for Warning Messages” on
page 83.
Solutions for Warning Messages
Error Code
Do this:
001, 002, 003, 004,
005, 006, 008, 032
Power off the control unit, then power it on
again.
009
1.
Make sure only the Hill-Rom approved
battery (194566S) is used.
2.
Press
return
after reading the on-screen
message.
3.
Make sure the battery is correctly installed in
the control unit. If necessary, remove the
battery and install it. Remove the battery and
install it. See “Install the Battery” on page 61“.
Draft_E_2019-Apr-29_Cleaned
Содержание Synclara PSC1
Страница 2: ...Draft_E_2019 Apr 29_Cleaned...
Страница 79: ...Synclara Cough System Model PSC1 User Manual 196286 REV E 71 Accessories Draft_E_2019 Apr 29_Cleaned...
Страница 109: ...Draft_E_2019 Apr 29_Cleaned...
Страница 110: ...Draft_E_2019 Apr 29_Cleaned...