18
Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Set up and Power on the Control Unit
•
Flexible tracheostomy adapter
(I)—
a.
Insert and gently twist the
tracheostomy adapter (I) into
the free end (D) of the
breathing hose.
b.
Connect the other end of the flexible tracheostomy adapter
to the tracheostomy tube (patient’s own).
SET UP AND POWER ON THE CONTROL UNIT
Warning
—Never block the air openings of the system or set it on a soft
surface, such as a bed or couch, where the air openings may be blocked.
Keep the air openings free of lint, hair, and the like. Otherwise, patient
injury could occur.
NOTES:
•
If the control unit is taken out from storage, do the following before
use:
–
Allow the control unit to
warm up
in an ambient
environment of 68°F (20°C) for at least
45 minutes
when
removed from a storage temperature of -4°F (-20°C) or
colder.
–
Allow the control unit to
cool down
in an ambient
environment of 68°F (20°C) for at least
80 minutes
when
removed from a storage temperature of 140°F (60°C) or
warmer.
•
If the control unit is on the cart, make sure the casters are set. For
instructions on mounting the control unit on the cart, see page 66.
1.
Set the control unit on a dry, flat, and hard surface. Make sure the
control unit is away from curtains, blankets, or any heat-generating
devices.
•
If a foot switch
1
is used, connect the foot
switch to the foot switch port on the
control unit. See “Foot Switch” on page 65.
1. The foot switch port is available only on custom configured systems.
Draft_E_2019-Apr-29_Cleaned
Содержание Synclara PSC1
Страница 2: ...Draft_E_2019 Apr 29_Cleaned...
Страница 79: ...Synclara Cough System Model PSC1 User Manual 196286 REV E 71 Accessories Draft_E_2019 Apr 29_Cleaned...
Страница 109: ...Draft_E_2019 Apr 29_Cleaned...
Страница 110: ...Draft_E_2019 Apr 29_Cleaned...