Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Cleaning and Disinfecting
•
Warning
—Clean the patient circuit between each therapy
session for the same patient.
•
Warning
—If the circuit and/or bio-filter are damaged or visibly
soiled, replace them. See “Replacement Parts/Kits” on page 72.
NOTE:
Each patient circuit is for use by a single patient and intended for 30 days
of treatment or a maximum of 90 treatment sessions.
These cleaning instructions are for multiple use of the circuit by a single
patient. The circuit has been tested for compatibility with mild detergent
and warm water for cleaning. The facility should follow their internal
infection control policies on cleaning the nebulizer and circuit. If a policy
is not in place, the following is a guideline for proper cleaning of the
circuit:
1.
Disconnect the circuit from the control unit.
2.
Disconnect the mouthpiece, face mask, adapters, handset, breathing
hose, and bio-filter.
3.
Examine the inside of the bio-filter for any damage and/or soilage. If
there is soilage or damage, replace the bio-filter immediately. For
examples of damaged bio-filters, see page 77.
4.
Wipe down the outside of the bio-filter with a compatible bacterial
cleaning solution. Do not use excessive liquid or harsh cleansers. See
“Compatible cleaning and disinfecting solutions” on page 74.
5.
Wash the mouthpiece or face mask, handset, and adapters with mild
detergent and warm water.
6.
Let the handset, mouthpiece, and adapters air dry or dry them with a
lint free cloth or paper towel.
7.
Wipe down the outside of the circuit tubing with an approved
alcohol-based cleaner.
8.
After all parts are dry, assemble the circuit, and put it in a breathable
bag with the patient’s name on it for use later.
NOTE:
Before the next use of the system, examine the components of the circuit
for cracks or other damage. If there is damage, replace the circuit. See
“Replacement Parts/Kits” on page 72.