Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
97
Specifications
Guidance and Manufacturer's Declaration—Electromagnetic
Immunity
The Synclara™ Cough System, Model PSC1 is intended for use in the electromagnetic
environment specified below. The customer or user of the Synclara™ Cough System,
Model PSC1 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 2 kV for
power supply
lines
Mains power quality should be that
of a typical home healthcare envi-
ronment.
± 1 kV for
input/ output
lines
Not
applicable
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 1 kV line(s)
to line(s)
Mains power quality should be that
of a typical home environment.
± 2 kV line(s) to
earth
Not
applicable
Voltage dips,
short
interruptions
and voltage
variations
on power
supply input
lines
IEC 61000-4-11
0% U
T
for 0.5
cycle at:
00, 450, 900,
1350, 1800,
2250, 2700, and
3150
0% U
T
for 1
cycle
70% U
T
; 25/30
cycles, single
phase at 00
0% U
T
for
250/300 cycles
0% U
T
for 0.5
cycle at:
00, 450, 900,
1350, 1800,
2250, 2700,
and 3150
0% U
T
for 1
cycle
70% U
T
; 25/30
cycles, single
phase at 00
0% U
T
for
250/300 cycle
Mains power quality should be of a
typical home healthcare environ-
ment. If it is necessary for the user
to have continued operation of the
Synclara™ Cough System during
power mains interruptions, it is rec-
ommended that the Synclara™
Cough System be powered from an
uninterruptible power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical home
healthcare environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Draft_E_2019-Apr-29_Cleaned