Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
61
Accessories
CAUTION:
Obey these
cautions
to help prevent equipment damage:
•
Caution
—Do not open, crush, puncture, incinerate, immerse in
water, or heat over 140°F (60°C) the battery.
•
Caution
—Use only the Hill-Rom approved battery (194566S).
•
Caution
—Fully charge the battery to 100% before the first time
use. Allow for up to 8 hours to fully charge the battery.
Subsequent battery recharge periods will take less time.
NOTE:
If the replaceable battery is unable to support at least 2 therapy sessions,
replace with a new and freshly charged battery. See “Replacement
Parts/Kits” on page 72.
For disposal of the battery, consult your local regulations to safely discard
or recycle the battery.
If you return the control unit for repairs, do not return the battery unless
specifically requested by customer service.
Install the Battery
NOTE:
These images show how to check that the battery is locked into position. If
you see an orange colored section in the latch area, then the battery is not
fully locked into position.
Press down on the battery until the orange colored section is no longer
visible.
1.
Turn the patient circuit hook
aside to get access to the
battery compartment.
2.
Remove the battery cover (if
installed).
3.
Insert the right side of the
battery (with LED indicator)
into the battery compartment
first.
4.
Insert the left side of the
battery until the battery locks
into position.
Draft_E_2019-Apr-29_Cleaned