Guidance and Manufacturer's Declaration—Electromagnetic
Immunity
The Synclara™ Cough System, Model PSC1 is intended for use in the electromagnetic
environment specified below. The customer or user of the Synclara™ Cough System,
Model PSC1 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the Synclara™
Cough System, including cables, than
the recommended separation dis-
tance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 V
150 kHz to
80 MHz
Radiated RF IEC
61000-4-3
10 V/m
80 MHz to 2.7
GHz
10 V/m
80 MHz to
800 MHz
800 MHz to
2.7 GHz
where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and
d
is the recom-
mended separation in meters (m).
Field strengths from fixed
RF transmitters, as deter-
mined by an electromag-
netic site survey
a
, should
be less than the compliance level in
each frequency range
b
. Interference
may occur in the vicinity of equipment
marked with the following symbol.
NOTES:
•
At 80 MHz and 800 MHz, the higher the frequency range applies.
•
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
•
Electromagnetic immunity was examined under normal operation at the time of testing. Normal
operation was the essential performance used. Electromagnetic immunity was verified at
settings of 4 for intensity and 12 Hz for frequency.
d
3.5
3
-------
P
=
d
3.5
10
-------
P
=
d
7
10
------
P
=
Draft_E_2019-Apr-29_Cleaned