Preventive Maintenance
Chapter 6: General Procedures
CareAssist™ Bed Service Manual (man330)
Page 6 - 7
6
CPR release
Inspect the handles, cables and CPR mechanism on the head motor.
Make sure the screws are present and fully tightened.
Make sure the two CPR release handles, cable and CPR mechanism
on the head motor are in good condition.
Raise the head section to the high position, then activate one of the
CPR releases. Make sure t the head section lowers. Adjust the CPR
cable as necessary. Do the same tests on the other side of the bed.
Release the CPR handles and make sure the mechanism locks
correctly by operating the Head Up control for a few seconds. The
elevation movement must be observed.
Replace the head section motor as necessary.
Head section gas
springs
Inspect the assembly of the gas springs. Check that the screws are
present and fully tightened.
Make sure there is no oil on its shaft.
Fully raise the head section. Activate the CPR function. Make sure
the head section lowers rapidly to the intermediate position then
gradually to the low position.
Replace as necessary.
Knee section motor
Inspect the knee section motor. Make sure the pins and retaining
clips are not missing.
Fully raise and lower knee section. make sure there is no friction or
abnormal noises and that no audible overload indication can be heard
during the movement.
Replace as necessary.
Automatic Contour
Make sure the knee section rises up to mid-height when the head
section is raised from the low position, and that it lowers
automatically when the head section lowers and the head section
reaches the low position.
Replace the actuator or defective unit if necessary.
Function
Procedure
Содержание CareAssist P1170
Страница 2: ......
Страница 6: ...Revisions Page iv CareAssist Bed Service Manual man330 NOTES ...
Страница 14: ...Page xii CareAssist Bed Service Manual man330 NOTES ...
Страница 141: ...Warranty Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 5 5 NOTES ...
Страница 144: ...Base Frame Chapter 5 Parts List Page 5 8 CareAssist Bed Service Manual man330 Base Frame Figure 5 2 Base Frame ...
Страница 147: ...Base Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 11 5 NOTES ...
Страница 151: ...Articulating Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 15 5 NOTES ...
Страница 157: ...Head Siderail Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 21 5 NOTES ...
Страница 165: ...CPR Cables Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 29 5 NOTES ...
Страница 170: ...Interface Control Box Chapter 5 Parts List Page 5 34 CareAssist Bed Service Manual man330 NOTES ...
Страница 187: ...SideCom Communication System Chapter 3 ...
Страница 188: ...Siderail Interface P C Board 72207 and 71514 Chapter 3 ...
Страница 189: ...Caregiver Control Chapter 3 ...
Страница 190: ...Left Caregiver Signal Conditioning Chapter 3 ...
Страница 191: ...Bed Control Board Power and Phase Control Chapter 3 ...
Страница 192: ...Bed Control Board Power Management Chapter 3 ...
Страница 193: ...Bed Control Board Battery Management Chapter 3 ...
Страница 194: ...Bed Control Board Clock and Divider Chapter 3 ...
Страница 195: ...Bed Control Board Relay FET and Drivers Chapter 3 ...
Страница 196: ...Bed Control Board PLD and Connector Chapter 3 ...
Страница 197: ...Bed Control Board Motor Current Limiter Chapter 3 ...