29
Oxygen Tank Holder, E-Size (P27601)
WARNING:
If the oxygen tank holder is placed at the foot end of the bed, ensure
the Knee Up/Down controls are locked out. Failure to do so can result
in caregiver, patient, or visitor injury if the foot section fully lowers and
the holder becomes dislodged from the bed.
CAUTION:
Do not exceed the load capacity of the oxygen tank holder. If the
oxygen tank is overloaded, personal injury or equipment damage may
occur.
The oxygen tank holder attaches to the head end of the articulating frame in a
vertical position. The oxygen tank holder accommodates one E size oxygen
tank with a regulator. The mounting points are located to allow the affixed oxygen tank holder to pivot.
To Install
•
Install the mounting bar vertically into a mounting socket at either the head end or foot end of the
articulating frame. Make sure the Knee Up/Down control is locked out if installing at the foot end.
•
Place the tank in the holder, and tighten the holder thumbscrew. The thumbscrew keeps the oxygen
tank from rotating in the holder.
To Remove
•
Loosen the thumbscrew that holds the tank secure in the holder.
•
Lift the tank out of the holder.
•
Lift up on the tank holder, and remove it from the mounting sockets.
Traction Frame Support (P1181)
The traction frame adapter bracket (P1181) is to attach fracture frame
equipment to the bed.
Refer to the equipment manufacturer’s instructions for installation procedures.
Patient Helper Support (P1180)
The Patient Helper Support (P1180) is used to attach the patient helper to the
bed.
Refer to the equipment manufacturer’s instructions for installation procedures