20
Bed Exit System (Beds without Scale)
WARNING:
The Bed Exit System is not intended as a substitute for good
nursing practices. The Bed Exit System must be used in
conjunction with a sound risk assessment and protocol.
The Bed Exit system is designed to be a reminder to the patient to
stay in bed. The Bed Exit control is located on the outside of both
head siderails.
To Activate
•
Ensure the patient is on the bed.
•
Simultaneously press both buttons on the Bed Exit control on either head siderail. The LED will
illuminate when armed.
When activated, the Bed Exit sounds a local alarm when the patient exits the bed. If the bed is equipped
with the SideCom® Communication System, and it is connected to the facility, a signal is sent to the
nurses station when the Bed Exit alarm sounds.
If the patient moves to exit the bed, the Bed Exit alarm sounds. If the Nurse Call system is operational, a
signal is sent to the nurse station. The Bed Exit system must be turned off at the bedside to cancel the
nurse call signal.
To Reset After the Alarm Sounds
•
Ensure the patient is on the bed.
•
Simultaneously press both buttons on the Bed Exit control on either head siderail to turn the system
off.
•
Simultaneously press both buttons on the Bed Exit control on either head siderail to turn the system
on. The LED will illuminate when armed.
To Deactivate
Simultaneously press both buttons on the Bed Exit control on either head siderail.
Bed Exit System (B Model Beds with Scale)
WARNING:
The Bed Exit System is not intended as a substitute for good nursing practices. The Bed Exit
System must be used in conjunction with a sound risk assessment and protocol.
The patient must weigh between 50 and 400 lb (23 to 181 kg) for Bed Exit to work.
To Activate
•
Press the Enable control on the scale control pod.
•
Press the Bed Exit control. When the Bed Exit control indicator comes
on, the Bed Exit is armed.
NOTE:
The Enable control on the scale control pod does not activate the Trendelenburg or Reverse
Trendelenburg controls.