Mattress Foot Pad Extender (P734EA1)
The mattress pad extender is used to fill in the gap between the mattress and the
footboard when the foot section is extended to its full length.
To Install
Insert the extender between the footboard and mattress after the foot section is
extended.
Infusion Support System Transfer Pole (P158)
WARNING:
If the ISS pole is placed at the foot end of the bed, ensure the Knee
Up/Down controls are locked out. Failure to do so can result in caregiver,
patient, or visitor injury if the foot section fully lowers and the ISS pole
becomes dislodged from the bed.
CAUTION:
Do not exceed the 20 lb (9 kg) load capacity of the ISS pole. If the ISS pole
is overloaded, personal injury or equipment damage may occur.
CAUTION:
Do not mount infusion pumps on the lower section of an IV pole. Doing so may cause
interference with head section articulation.
CAUTION:
When lowering the upper section of an IV pole, always grasp and hold the upper section of the
pole before pulling the release knob.
The Infusion Support System (ISS) consists of a movable and adjustable IV pole. The pole supports IV
pumps or bags in a vertical orientation and raises or lowers the pumps or bags with respect to the system
frame. The ISS pole can hold 20 lb (9 kg).
The head end of the system has attaching points for two mobile ISS. Each ISS can support one infusion
pump plus two liters of intravenous solution.
When using the ISS, it is necessary to use the P163 socket adapter before installing the ISS pole.
Mattress (P731EA3)
WARNING:
Only use mattresses of the specified dimension. Failure to do so could result in personal injury.
The mattress contains an all-foam, zoned, three-layered core with foam side bolsters and is designed to
reduce patient interface pressures. The CareAssist™ Bed surface is designed especially for the
CareAssist™ Bed frame. Loose fitting sheets (preferably knitted) must be used for correct operation of
the sleep surface.
The CareAssist™ Bed surface is designed especially to work with the following system features:
•
Shearless Pivot® Patient Position Mechanism
•
FullChair® Patient Position Mechanism
Loose fitting sheets (preferably knitted) must be used for correct operation of the sleep surface.