H i k o n e b | 9 0 8 D C
16
User Manual
Note
• Due to its chemical characteristics, the nebulization chamber, medication cup
and aerosol & ventilation tubes are not suitable for ether type lubricants, ketones,
hydrocarbons and cyclohexanone type substances.
Warning
Low Level Water alarm: If the water in the Nebulization Chamber falls below the lower
limit, the warning led will illuminate and the unit will alarm. When water is added, both
warning LED and alarm will deactivate.
8. Troubleshooting
Error
Reason
Solution
The device is not
operational,
On-Off button light is
not on
No mains
Check plug, IEC connections and
fuses
On-Off button is on
but there is no aerosol
production
1. Transducer
2. O/C fuse
1.Replace transducer
2. Replace fuses
Red error LED is on
Low Water Level
Nebulizing chamber not locked
on the device correctly
1. Add more water
2. Re-fit chamber ensuring correct
Aerosol production is
visible, but not coming
out of tubing
Blockage of patient tube due to
wrong positioning
Occluded Filter
Flow menu disabled
Faulty fan
Change position of the aerosol tube
to point upwards and straight
Replace the filter
Activate Flow setting
Replace Fan