19
HN 908 DC_EN Rev. A
H i k o n e b | 9 0 8 D C
User Manual
Disinfection
When disinfecting the device surface, the device’s surface, make sure that the mains
power cable is removed. Ensure that no liquid enters inside the device when cleaning.
Operate the device only after the disinfection has completely dried and evaporated. Use
only disinfection products consisting of aldehyde, ammonia and alcohol components.
Never use phenol derivative disinfections, as this will compromise and reduce the
service life of plastic parts.
Examples of surface disinfectants:
- Incidin® Plus *
- Incidin® Perfect *
- Antiseptica Kombi surface disinfectant *
- Antifect® FF *
- B 10 *
Examples of surface disinfectants:
- Cidex® AF
- Gigasept® FF *
* Registered trademarks of the respective manufacturers
The lifespan of the nebulizer chamber with lid and the quartz module is limited to 30
autoclaving cycles. The example disinfectants are recommendations from the device
manufacturer. The recommendation of the mentioned active ingredient bases (pH 4.5-
8) applies. This does not release the user from the need to consult a hygiene specialist
in individual cases or to test the components to be disinfected for compatibility with the
active ingredients.
Содержание 908 DC
Страница 1: ...EN DE ...
Страница 11: ...11 HN 908 DC_EN Rev A H i k o n e b 9 0 8 D C User Manual 14 15 19 36 38 11 22 8 16 23 21 6 3 26 1 5 4 ...
Страница 25: ...25 HN 908 DC_EN Rev A H i k o n e b 9 0 8 D C User Manual ...
Страница 50: ......