Each command must be ended with ASCII code <CR>.
?
Request for displayed value. Transmission format as described in chapter 11.1.
FNC x
Choice of the measuring function
X=
0
1
2
3
4
5
6
7
8
Func=
—
A
-A
B
-B
A+B
A-B
-A+B
-A-B
FNC?
Request for the measuring function
ID?
Request for serial number of TE TT80, TE TT90 or TE UPC
MM / IN
inch/metric conversion
MES?
Request for the measurement-type, internal or external (INT, EXT)
PRE xxx
Input of offset value for active function
xxx = true (x.xxxxx or -x.xxxxx in mm / x.xxxxxx or -x.xxxxxx en inch)
PRZ
Zero-setting of offset value for each active measuring channel
RNG x
Choice of the measuring range
X=
0
1
2
3
4
5
6
7
8
9
Range µm
±5000
±2000
±500
±200
±50
±20
±5
±2
±0.5
Auto
Range in
±.200
±.100
±.020
±.010
±.002
±.001
±.0002
±.0001
±.00002
Auto
TT80
Range in/µin
±.100
±.020
±.010
±5000
±2000
±1000
±200
±100
±20
Auto
TT90
RNG?
Request for the used range
RST
Resets the instrument to its defaulted configuration (Reset)
STO0, STO1
Enables/Disables the measured value (HOLD function)
TOL xxx xxx
Input of tolerance limits
xxx = true (x.xxxxx and -x.xxxxx in mm / x.xxxxxx and -x.xxxxxx in inch)
TOL?
Request for tolerance limits
TO0, TO1
Enables/Disables the tolerance mode (in the TT90-UPC mode,
this command is ignored and the error message ERR23 displayed).
UNI?
Request for actual unit : TT8 = mm / in – TT90 = µm / mil
VER?
Request for the programme version
MEM x
Selection of the measuring modes Min, Max, Max-Min or (Max+Min)/2
X =
0
1
2
3
4
Func. MEM= Normal
Max
Min
Max-Min (Max+Min)/2
MEM?
Request for the memory function
STM
Sets the memories to actual value
30