9.7 Analogue output
The analogue output is linked to the analogue indication for the measuring ranges of
± 5000 µm/± 0.020 in. Therefore, the voltage in full scale matches ± 2V/±10V.
For ranges of ± 200 µm/± 0.010 in, this output is the only one available with a voltage to
± 2V/±10V for a maximum value of ± 200 µm/0.010 in.
Ranges µm
Voltage in full scale
Sensitivity mV / µm
± 5000
±2 V
±10 V
0,4
2
± 2000
±2 V
±10 V
1
5
± 500
±2 V
±10 V
4
20
± 200
±2 V
±10 V
10
50
±
50
±0,5 V
± 2,5 V
10
50
±
20
±0,2 V
± 1 V
10
50
±
5
±0,05 V
± 0,25 V
10
50
±
2
±0,02 V
± 0,10 V
10
50
±
0.5
±0,005 V
± 0,025 V
10
50
Maximum load capacitance: 40nF, max. 3mA.
For inch ranges, refer to chapter 6.3.