12
RANGE
RANGE
1
± 5000
µm
200
µm
± 0.200
in
0.010
in
2
± 2000
µm
100
µm
± 0.100
in
0.005
in
3
±
500
µm
20
µm
± 0.020
in
0.001
in
4
±
200
µm
10
µm
± 0.010
in
0.0005
in
5
±
50
µm
2
µm
± 0.002
in
0.0001
in
6
±
20
µm
1
µm
± 0.001
in
0.00005 in
7
±
5
µm
0,2 µm
± 0.0002
in
0.00001 in
8
±
2
µm
0,1 µm
± 0.0001
in
0.000005 in
9
±
0,5 µm
0,02 µm
± 0.00002 in
0.000001 in
10
Auto Range
Auto Range
6.3
Analogue measuring ranges
TESATRONIC TT80 and TT90 feature 9 switchable measuring ranges plus one built-in auto
range.
Each pressure on this key causes the selected range to change.
Ranges
Metric range
Scale interval
Inch range
Scale interval
1
± 5000
µm
100
µm
± 0.100 in
0.005
in
2
± 2000
µm
20
µm
± 0.020 in
0.001
in
3
±
500
µm
10
µm
± 0.010 in
0.0005 in
4
±
200
µm
10
µm
± 5000 µin
200 µin
5
±
50
µm
2
µm
± 2000 µin
100 µin
6
±
20
µm
1
µm
± 1000 µin
50 µin
7
±
5
µm
0,2 µm
±
200 µin
10 µin
8
±
2
µm
0,1 µm
±
100 µin
5 µin
9
±
0,5 µm
0,02 µm
±
20 µin
1 µin
10
Auto Range
Auto Range
Ranges
Metric range
Scale interval
Inch range
Scale interval
The measuring range of both TT90 and TT90-UPC is restricted to ±200 µm/0.008 in.
However, ranges from 500 up to 5000 µm (0,02 à 0.2 in) may additionally be used for dis-
playing the measured value. In the chart below, there’s no relationship between the right-
hand (µm) and left-hand (in) columns.
With display set to µin, the numerical value can be as high as 8000 µin.
TT90 and TT90-UPC
Содержание 04430011
Страница 2: ...TESATRONIC TT80 TT90 Instruments lectroniques de mesure de longueurs Mode d emploi Fran ais...
Страница 34: ...TESATRONIC TT80 TT90 Elektronische L ngenmessger te Gebrauchsanleitung Deutsch...
Страница 66: ...TESATRONIC TT80 TT90 Electronic length measuring instruments Instruction manual English...