EN
AB1101EN_SA
Rev. 01 / 11.2022
33/45
20.9
Resetting the cycle counter to zero
After the rotor/hanger has been exchanged, the cycle counter of the respective rotor code must be reset to zero
again.
The cycle counter may only be reset to zero if the rotor/hanger has been exchanged for a new rotor/hanger
first.
With the rotor at a standstill, the cycle counter can be reset as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
Press and hold the button for eight
seconds.
PROG
t/min:s
RPM
Press the key until the following is
displayed:
PROG
t/min:s
RPM
START
PULSE
Press the key.
Example:
PROG
t/min:s
RPM
Press the key until the following is
displayed:
Cyc sum :
Number of completed run
cycles.
PROG
t/min:s
RPM
RCF
Press the key.
PROG
t/min:s
RPM
Press the key.
The number of completed run cycles is
reset to zero.
PROG
t/min:s
RPM
START
PULSE
Press the key to save the setting.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
STOP
OPEN
Press the key twice to exit the "
Time &
Cycles" menu or press three times to exit
the " MACHINE MENU ".
PROG
t/min:s
RPM
Example:
PROG
t/min:s
RPM