EN
32/45
Rev. 01 / 11.2022
AB1101EN_SA
20.8
Querying the hours of operation and the number of centrifugation runs
The hours of operation are divided into internal and external hours of operation.
Internal hours of operation: Total time the device was switched on.
External hours of operation: Total time of the previous centrifugation runs.
With the rotor at a standstill, the query can proceed as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key.
Press and hold the button for eight
seconds.
PROG
t/min:s
RPM
Press the key until the following is
displayed:
PROG
t/min:s
RPM
START
PULSE
Press the key.
TimeExt :
External hours of operation.
Example:
PROG
t/min:s
RPM
Press the key.
TimeInt :
Internal hours of operation.
Example:
PROG
t/min:s
RPM
Press the key.
Starts :
Number of all centrifugation runs.
Example:
PROG
t/min:s
RPM
STOP
OPEN
Press the key twice to exit the "
Time &
Cycles" menu or press three times to exit
the " MACHINE MENU ".
PROG
t/min:s
RPM
Example:
PROG
t/min:s
RPM