20.4
Automatic unlocking of the lid after the centrifugation run
It can be set whether the lid should automatically unlock or not after the centrifugation run.
With the rotor at a standstill, this can be set as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
Press and hold the button for eight
seconds.
PROG
t/min:s
RPM
Press the key until the following is
displayed:
PROG
t/min:s
RPM
START
PULSE
Press the key.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
Press the key until one of the following is
displayed:
Lid AutoOpen :
Automatic unlocking of
the lid after the centrifugation run.
off :
Lid does not unlock automatically.
on :
Lid unlocks automatically.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
With the keys, set
off
or
on
.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
START
PULSE
Press the key to save the setting.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
STOP
OPEN
Press the key once to exit the "
Settings"
menu or press twice to exit the "
MACHINE MENU ".
PROG
t/min:s
RPM
Example:
PROG