Important: Always read and follow the instructions.
• 17
SELF-DIAGNOSTICS
The machine can perform self-diagnostics to check the LED’s on the control panel and make sure the encoder
reads correctly.
To perform this operation, view the SETUP menu.
In the self-diagnostics sequence, all the LED’s on the panel light up for a few seconds in order to check
operation. When the LED’s go out, the machine automatically moves on to the encoder reading phase. When the
wheel is turned manually (forwards and backwards), the display shows its exact position. The value lies between 0
(zero) and 255.
(Consult the maintenance manual or contact Technical Service).
BALANCING MACHINE CALIBRATION
(Calibrate in OZ mode)
To calibrate the balancing machine, use a wheel with steel rim of average dimensions,
(
e.g.
1
6”).
To properly perform the procedure:
▪
Mount a wheel on the machine, even unbalanced, and very carefully set its dimensions.
Setting incorrect measurements will result in the machine not being correctly calibrated, and balancing of subsequent wheels
will hence be incorrect until the machine is recalibrated with the correct measurements!!
If Err. 241 occurs, manually set the width value before accessing the wheel balancer autocalibration function.
1. Display the CALIBRATION function from the SETUP menu
Close the guard and press the
button to perform a spin.
2.
Add a standard weight of
4.00 oz
anywhere on
the outer side
plane.
Close the guard and press the
button to perform a spin.