7-6 Initialization of the Scale to Printer (Optional)
Step 1
Press
G N/SET
key while powering on or long press
G N/SET
key in the
weighing mode to enter parameter setting. Press key
e
/MR
or
TARE/
f
to circle
the parameters and select
SET08
。
Step 2
Press
G N/SET
to enter the external equipment choosing. Press key
e
/MR
or
TARE/
f
to select the type of printers. Press
G N/SET
to start initialization.
The window of unit weight displays the type of printer. And the window of total
amount displays UNSU or init
?。
Step 3
When UNSUP appears, it means the printer can’t be initialized. Press key
ZERO/ESC
to return to the printer choosing. It needs to be initialized when init?
appears. Press key
M+/SAVE
to start initialization or press key
ZERO/ESC
to
return to the printer choosing without initialization. It displays ok when initialization
is finished and shows the type of printer after one second. Go further to other
parameter setting or press key
ZERO/ESC
to return to weighing mode.
7-7.The Control from PC to Scale (Optional)
Step 1
Connect the indicator and computer.
Step 2
Open the serial port to modify. Input the letters Z, T, C , R and P in the sending
window and press key Send again. The scale can conduct the responding
function. Z= Zero, T=Tare, R and P=Printer, C=Cancel Tare.
8. Serial Interface
If external interface is needed, please select the proper RS 232 board firstly, and only when
this board is adopted, the functions can be enabled.
Note:
RTC is unavailable.