9
1. Press and hold
TARE/
while powering on.
Do not release it till the window displays
“
CAL”
2. With no load on the weighing pan,
press
TARE/
to start zero point calibration.
3. Wait till the window displays the first calibration value.
Note:
The first calibration value is default. If the value
has been changed, the default value is 1/3 of full load.
Note:
If you need to change the value, do as the following: Press key
G N/SET
to enter
the value setting. Press
key
e
/MR
or
TARE/
f
to move leftwards or rightwards. Press
key
M-/HI
or
PRINT/LO
to modify the value (or via numeric keys). Press key
M+/SAVE
to save.
4. Put the corresponding weight on the weighing pan,
and then press
TARE/
to complete the first point calibration.
Note:
After the first point calibration, the window can display the weight value. If no need
for the other point calibration, move to Step 6 to finish the calibration procedure.
5. Add another weight 500g to the current weight. The window
will show the total weights value. Press
TARE/
to complete.
Repeat this step to complete the calibrations of on3, on4 and on5.
6. Press
ZERO/ESC
to save. After the window displays “
”,
It will return to normal weighing mode.
Note
: the window displays
automatically when on5 is finished. It will be saved and
return to the normal weighing.
7. Operation
7-1 Weighing
Put item to be weighed on the scale. The Weight window shows 1.00kg (Gross Weight).
7-2 Tare & Preset Tare
Tare
When weighing a sample that must be held in a container, tare and store the container weight
Содержание CYD-C8+-15K
Страница 22: ...21 ...