11
2.13
Adjustment
1.
Turn the unit on.
2.
Pull knob (1) on water filter up and adjust the water pressure to 0.5
Bar.
3.
Activate the syringe with water until all air has left the system.
4.
Adjust the water pressure to 2 Bar and push button (1) down again.
5.
Pull bottom (2) on the air filter up and adjust air pressure to 4.5 Bar.
6.
Mount the spittoon rinse pipe (3) and the cup fill pipe (4).
7.
Place a cup in the cup holder. Program the cup fill – and the spittoon
rinse time.
Содержание UnicLine 5D
Страница 7: ...7...
Страница 13: ...13 3 Final Test To be completed by authorized technician...
Страница 15: ...15 To be completed by authorized technician and the Dentist...
Страница 16: ...16 5 Installation plan and floor design 500 min 440 min 440 min 540 465 735 500...
Страница 18: ...18 5 1 Drilling diagram Please use the enclosed diagram when placing the unit in the clinic...
Страница 19: ...Baldersh j 38 DK 2635 Ish j Denmark Tel 45 4332 0990 Fax 45 4332 0980 www heka dental dk...