9
2.9
Patient chair
1.
Unwrap the chair.
2.
Gently place the chair on the side, leaning against the floor. Mount the
foot control. Connect the socket to the distribution board.
3.
Gently place the chair in its normal position again. Mount the chair on
the floor as shown in the installation plan.
4.
Mount the armrest. (If ordered)
5.
The Cable from the chair leads through the “relief” connecting piece.
(page 7) Connect it to the power distribution board terminal, labelled:
Chair.
6.
Mount the cushion. Push the keyhole cuts on the backside of the
cushion, gently over the screw heads. (Begin on the end with leg
support.)
7.
Mount the back cushion.
8.
Place the UnicLine sticker (5).