2.10
Connections
1.
Rinse the water supply and ventilate the air supply completely, before
connecting the unit (for at least for 5 min.).
2.
Connect the green tube (water) to the tap. Use the supplied fittings
and the water filter input slot. Open the water supply. Make sure you
tighten and inspect all connections for any leaks.
3.
Connect the blue tube (air) to the tap. Use the supplied fittings and
the air filter input slot. Open the air supply. Make sure you tighten
and inspect all connections for any leaks.
4.
Connect the suction system from the unit to the suction system in the
Clinic. The Unit can be delivered with either wet- or dry suction.
5.
Connect the waste tube from the spittoon and cup to the supplied
fittings. Connect the air filter ventilation tube.
6.
Connect the waste fitting to the clinic waste pipe.
7.
Mount the suction motor cable to the power distribution board
terminal labelled ‘Aspir’.
8.
Connect main power to the power distribution board terminal labelled
‘Mains’.