background image

 

 

 

2. 

Installation 

2.1 

Static electricity 

Avoid electrostatic discharge! 

Before touching any of the electronics, make sure the equipment is switched off 
and the unit is connected with the equipment’s ground. 
 
In an emergency, hold on to the console during the mounting. 
 

2.2 

Installation requirements 

For installation placement, see Chapter 5 – Installation plan and floor design 
 
During installation, it is the responsibility of the technician to ensure that all 
installations meet and is implemented in compliance with national and 
international requirements.  
 

1.

 

230 Volt +/- 10%. 50 Hz with ground. Main fuse 10A. The cable must have 
a free 

end of at least 35 cm. 

2.

 

Water, pressure 2 - 6 bar. Balltap with R 3/8” inner thread. Top 5 cm 
above the floor. 

3.

 

The water should be drinking quality and if the water has hardness greater 
than 8 ºdH (1 ºdH = 20 mg calsium/3 l water) a water conditioner must be 
fitted to the water inlet pipe. Hard water can very quickly ruin a dental 
unit. 

4.

 

Air, pressure 4,5-8 bar. Balltap with R 3/8” inner thread. Top 5 cm above 
the floor. 

5.

 

Drain, outside diameter 35 mm, PVC tube, Top 5 cm above the floor 
(DN40). 

6.

 

Suction, outside diameter 32 mm, PVC tube, Top 5 cm above the floor 
(DN40) 

7.

 

Steering wire to suction motor, 2x1,5mm^2, Free end 50 cm above the 
floor. 

8.

 

Steering wire to call systen, 2x1,5 mm^2, Free end 50 cm above the floor. 

9.

 

For wiring to X-ray, the type and number of cable depends on the 
manufacture. Free end minimum is 50cm above the floor. 

 
Installation of the cable can be made over the floor. If this is the case, drill a hole 
in one of the side parts. The equipment must not be mounted near strong radio- 
or radar installations. 
 

2.3 

Console 

1.

 

Drill 6 holes Ø 10 mm in the floor according to the drill plan or mounting 
pattern. Chapter 5.1 

2.

 

Tighten 2 nuts on each mounting bolt and screw the bolts into the drilled 
holes. If the floor is made of a porous material, request a floor mounting 
plate (item no. a-482(9010)) with the unit. 

Содержание UnicLine 5D

Страница 1: ...3 Mounting instructions UnicLine 5D UnicLine patient chair Dental light KEEP THIS MANUAL WITH UNIT AT ALL TIMES Manuals for OEM equipment are included in the shipping boxes Installation Service and M...

Страница 2: ...3 Console 4 2 4 Instrumentarm and instrument table 6 2 5 Foot control 8 2 6 Operating lamp 8 2 7 X ray machine 8 2 8 Call system 8 2 9 Patient chair 9 2 10 Connections 10 2 13 Adjustment 11 3 Final Te...

Страница 3: ...ng parts UnicLine patient chair Dental light Unpack the unit and chair with caution Remove the bubble plastic around the unit console and demount the sides of the console by turning the 2 large screws...

Страница 4: ...r must be fitted to the water inlet pipe Hard water can very quickly ruin a dental unit 4 Air pressure 4 5 8 bar Balltap with R 3 8 inner thread Top 5 cm above the floor 5 Drain outside diameter 35 mm...

Страница 5: ...a level on the arm s contact face on the console where the arm set is to be mounted and adjust the balance screws with the enclosed tool until the console is levelled The console is to be as close to...

Страница 6: ...istribution device 6 The blue hose air is mounted on the air distribution device 7 5 Mount the Return air device 7a and connect the yellow hose Dim 4 between 7b and 7c 6 Attach the yellow return air h...

Страница 7: ...7...

Страница 8: ...power distribution board terminal labelled Lamps If the unit is provided with a turn on turn off function please connect the lamp cable to the circuit labelled ON OFF 2 7 X ray machine 1 Mount the ad...

Страница 9: ...unt the chair on the floor as shown in the installation plan 4 Mount the armrest If ordered 5 The Cable from the chair leads through the relief connecting piece page 7 Connect it to the power distribu...

Страница 10: ...and the air filter input slot Open the air supply Make sure you tighten and inspect all connections for any leaks 4 Connect the suction system from the unit to the suction system in the Clinic The Un...

Страница 11: ...yringe with water until all air has left the system 4 Adjust the water pressure to 2 Bar and push button 1 down again 5 Pull bottom 2 on the air filter up and adjust air pressure to 4 5 Bar 6 Mount th...

Страница 12: ...the dentist s desired quantity using the knobs under the instrument table 11 Position any hand held instruments the dentist normally uses together with the surgery equipment 12 Dismount cover on the b...

Страница 13: ...13 3 Final Test To be completed by authorized technician...

Страница 14: ...plate for the console Attention The installation card must be completed and returned to Heka Dental A S by an authorized service engineer The two annual service cards must be completed by an authoriz...

Страница 15: ...15 To be completed by authorized technician and the Dentist...

Страница 16: ...16 5 Installation plan and floor design 500 min 440 min 440 min 540 465 735 500...

Страница 17: ...leitung f r Saugmaschine 5x1 5 Conducteur pour pompe aspirante 2x1 5 Styreledning til sug 2x1 5 F Water outflow without thread 32 mm Wasserabfluss ohne Gewinde DN40 Evacuation d eau sans filetage 32 m...

Страница 18: ...18 5 1 Drilling diagram Please use the enclosed diagram when placing the unit in the clinic...

Страница 19: ...Baldersh j 38 DK 2635 Ish j Denmark Tel 45 4332 0990 Fax 45 4332 0980 www heka dental dk...

Отзывы: