T32FNA
-
The amplitudes (peak-to-peak) must never exceed 70 % of the nominal
torque for the specific type, even with oscillating torque. In all cases the
amplitudes must lie within the load range limits of -- M
N
and + M
N
.
+M
N
100
-M
N
100
50
50
0
45
25
70% M
N
Range
Nominal torque M
N
in %
Upper torque limit
Lower torque limit
Fig. 6.1:
Amplitudes of dynamic torque
7
Maintenance
Torque transducers type T32 FNA are maintenance-free owing to their design
without bearings and with non-contacting transfer of the energising voltages
and of the measured signal.