![HBM CX27B Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759041.webp)
Mechanical
Quantum
X
I3031-14.0
HBM: public
41
6
Mechanical
QuantumX modules are extensively tested. This includes
- the extended temperature range
- mechanical vibration with an amplitude of 50 m/s²
in the frequency range 5 ... 2000 Hz in all 3 axes
for 2 hours, and
- the effect of exposure to 1000-fold mechanical
shock with an acceleration (half cosine) of 350 m/s²
for 3 ms in all 3 axes.
The degree of protection given in the technical data indic
ates the suitability of the housings for various ambient
conditions and also the protection of persons against po
tential risks when used. The letters
IP
(International Pro
tection), which are always present in the designation, are
followed by two digits. These indicate which degree of
protection a housing offers against contact or foreign
bodies (first digit) and moisture (second digit).
QuantumX modules are in a housing with IP20 as degree
of protection.
IP
2
0
Code
index
Degree of protection against
contact and foreign bodies
Code
index
Degree of protection against
water
2
Protection against contact with
fingers, protection against for
eign substances with
>12
mm
0
No water protection
Both housing types can be connected together with the
aid of two lateral housing clips (1-CASECLIP, not in