![HBM CX27B Скачать руководство пользователя страница 131](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759131.webp)
Modules and transducers
Quantum
X
I3031-14.0
HBM: public
131
System LED
Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs
All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing
orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running
(calibration)
Green
Error-free operation
Flashing green (5s),
then green
TEDS data being read in
Flashing orange (5 s),
then green
Manual configuration ongoing (ignore TEDS)
Red
Amplifier overload, no sensor connected
Channel error (incorrectly parameterized, connection error,
invalid TEDS data)
Flashing red
Sensor supply overload
General rule:
Brief flashing
TEDS identified (green: is used,
orange: is not used).